Name: TEST KIT, VERIFY BOWIE DICK (20/BX)
• Contains Latex
• Consists of a series of steam penetration (air removal) barriers in the center of which is a chemical indicator sheet
• The indicator sheet may be stored as a permanent record of the test.
• Used to qualify newly installed prevacuum sterilizers or following a major repair
• Designed to monitor performance changes in air removal and steam quality during routine use
• Equivalent to AAMI Bowie-Dick Test Pack